Analisis Nilai CTDIvol Dan DLP Pada Pemeriksaan CT Scan Brain Non Kontras Dewasa Untuk Periode Januari – Desember 2022 Di Instalasi Radiologi Rumah Sakit X Jakarta Pusat

Authors

  • Pricilia Shalry Horuoby Akademik Teknik Radiodiagnostik dan Radioterapoi Bali, Indonesia
  • Anak Agung Aris Diartama Akademik Teknik Radiodiagnostik dan Radioterapoi Bali, Indonesia
  • I Kadek Sukadana Akademik Teknik Radiodiagnostik dan Radioterapoi Bali, Indonesia

DOI:

https://doi.org/10.55606/jrik.v3i3.2687

Keywords:

CTDIvol, DLP, CT Brain Scan

Abstract

Background: CT scans are used for various types of examinations such as examinations of the head, thorax, abdomen and so on. The most frequently performed CT scan is a CT scan of the head. However, in the medical field, especially radiology, CT scans have the highest dose compared to other radiation equipment. Efforts to reduce radiation doses and radiation effects are by estimating the correct dose given to CT Scan patients using an optimization index called DRL (Diagnostic Level Reference), namely by evaluating the values ​​of CTDIvol and DLP on CT scans of the head, which have been determined in accordance with BAPETEN/I-DRL 2021 standards.

Method: This research is a descriptive quantitative study with an observational approach which aims to analyze CTDIvol and DLP values ​​based on retrospective data on non-contrast brain CT scans at the Radiology Installation at Hospital X, Central Jakarta.

Results: Calculation of the 2nd quartile (50th percentile) value of CTDIvol and DLP in Adult Non-Contrast Brain CT Scan examinations for the January-December 2022 Period at the Radiology Installation at X Hospital, Central Jakarta is CTDIvol 55.51 mGy and DLP 867.00 mGy* cm.

Conclusion: The 50 percentile value of CTDIvol and DLP in the Adult Non-Contrast Brain CT Scan examination for the January-December 2022 period at the Radiology Installation at X Hospital, Central Jakarta is in accordance with the standards set by BAPETEN/I-DRL 2021. Where is the value for CT Non-Contrast Head Scan, namely a CTDIvol value of 60 mGy and a DLP value of 1275 mGy*cm.

References

AJ T. Dose Assessment in Computed Tomography Examination And Estabilishment Of Local Diagnostic Reference Levels In Mazandaran, Iran. Natl Libr Medice. 2015;84–177.

Badan Pengawasan Tenaga Nuklir (BAPETEN). Rekaman Dokumen S. Pedoman Teknis Penyusunan Tingkat Panduan Diagnostik Atau Diagnostic Reference Level (Drl) Nasional. Jakarta. 2016;(8):63858275.

Badan Pengawas Tenaga Nuklir (BAPETEN). Keputusan Kepala Badan Pengawas Tenaga Nuklir Nomor: 1211/K/V/2021 Tentang Penetapan Nilai Tingkat Panduan Diagnostik Indonesia (Indonesian Diagnostic Reference Level) Untuk Modalitas CT-Scan Dan Radiografi Umum. 2021;4.

Cory Amelia ES dan ZA. Menggunakan Ct Dose Profiler Phantom. Youngster Phys J. 2014;3(4):329–34.

Diartama, AA Aris., Lobang Valey J., Wirajaya, I Wayan A., & Pratista IMS. Evaluasi Dosis Computed Tomography Dose Index Volume (CTDIvol) Dan Dose Length Product (DLP) Pada Pemeriksaan Multislice Computerized Tomography (MSCT) Thorax Dewasa Di RSU X Denpasar. J Ilmu Kedokt Dan Kesehat. 2023;10, No.4

Gede Agus Krisna Yogantara P, Ngurah Sutapa G, Made Yuliara I, Studi Fisika P, Matematika dan Ilmu Pengetahuan Alam F. Analisis Dosis Efektif Pada Pemeriksaan Computer Tomography (CT) Scan Kepala Di RSUD Sanjiwani Gianyar Effective Dose Analysis on Computer Tomography (CT) Head Scan at Gianyar Sanjiwani Hospital. Akreditasi SINTA 4 Mulai. 2021;22(2):53–9.

Irsal M, Winarno G. Pengaruh Parameter Milliampere-Second (mAs) terhadap Kualitas Citra Dan Dosis Radiasi Pada Pemeriksaan CT scan Kepala Pediatrik. J Fis Flux J Ilm Fis FMIPA Univ Lambung Mangkurat. 2020;17(1):1.

Johnson PT, Fishman EK. Enhancing Image Quality in the Era of Radiation Dose Reduction: Postprocessing Techniques for Body CT. J Am Coll Radiol [Internet]. 2018;15(3):486–8. Available from: https://doi.org/10.1016/j.jacr.2017.11.001

Lampignano JP. Radiographic Positioning And Related Anatomy. 7th ed. 2018. 9 p.

Masuku RS. Nautical : Jurnal Ilmiah Multidisiplin Analisis nilai Computer Tomography Dose Index ( CTDI ) dan Dose Length Product ( DLP ) pada pemeriksaan Multislice Computed Tomography ( MSCT ) trauma kepala dewasa di instalasi radiologi Rumah Sakit Balimed Denpasar. 2023;1(10):1235–43.

Pearce EC. Human Anatomy And Physiology. Indonesia U, editor. Jakarta: Gramedia Pustaka; 2016.

Pratista S. Estimasi nilai dosis radiasi pada pemeriksaan CT-Scan Kepala Dewasa Non Kontras berdasarkan nilai CTDI dan DLP di Instalasi Radiologi RS TK II Pelamonia. 2023;1(10):1147–52.

Rasad PHS. Radiologi Diagnostik. Ekayuda I, editor. Jakarta : Divisi Radiodiagnostik, Departemen Radiologi.Fakultas Kedokteran Universitas Indonesia; 2018.

Rumah D, Tk S, Pelamonia II, Pratiwi NJ, Agung A, Diartama A, et al. Garuda Pelamonia Jurnal Keperawatan Evaluasi Nilai CTDI VOL Dan DLP Pada Pemeriksaan MSCT Abdomen Non Kontras Selama Periode Maret 2022 Sampai Juni 2022 Program Studi Sarjana Terapan , Teknilogi Radiologi Pencitraan ATRO Bali Garuda Pelamonia Jurnal Keper. Garuda Pelamonia J Keperawatan. 2022;4(2):37–50.

Seibert AJ. “The Standardized Exposure Index For Digital Radiography: An Opportunity For Optimization Of Radiation Dose To The Pediatric Population” Pediatr. Radiol.,. 2014;41:573–81.

Sri WNN, Juliantara IP k, Darmita MP. Estimasi Dosis Radiasi Pada Pemeriksaan Ct Scan Kepala. J Radiogr Indones. 2018;107–10.

Wahyuni S, Amalia L. Perkembangan Dan Prinsip Kerja Computed Tomography (CT Scan). Galen J Kedokt dan Kesehat Mhs Malikussaleh. 2022;1(2):88.

Wibisino N. Koreksi Geometri Pengukuran Dosis pada Phantom menggunakan Metode CTDI. 2011;

Downloads

Published

2023-11-20

How to Cite

Pricilia Shalry Horuoby, Anak Agung Aris Diartama, & I Kadek Sukadana. (2023). Analisis Nilai CTDIvol Dan DLP Pada Pemeriksaan CT Scan Brain Non Kontras Dewasa Untuk Periode Januari – Desember 2022 Di Instalasi Radiologi Rumah Sakit X Jakarta Pusat. Jurnal Rumpun Ilmu Kesehatan, 3(3), 280–293. https://doi.org/10.55606/jrik.v3i3.2687